Sinuo Testing Equipment Co. , Limited

Sinuo Testing Equipment Co., Limited

Manufacturer from China
Verified Supplier
4 Years
Home / Products / Environmental Test Chamber /

Environmental Testing Constant Temperature Test Chamber IEC60068-2-2 512L

Contact Now
Sinuo Testing Equipment Co. , Limited
Visit Website
City:guangzhou
Province/State:guangdong
Country/Region:china
Contact Person:MsZoe Zou
Contact Now

Environmental Testing Constant Temperature Test Chamber IEC60068-2-2 512L

Ask Latest Price
Video Channel
Brand Name :Sinuo
Model Number :SN889-512L
Certification :Calibration Certificate (Cost Additional)
Place of Origin :China
MOQ :1
Price :Customized
Payment Terms :T/T
Supply Ability :1 sets/month
Delivery Time :45 days
Packaging Details :Other
Inner chamber dimensions :800*800*800 mm (W*H*D)
Outer chamber dimensions :1080*1950*1800 mm (W*H*D)
Temperature Range :20 °C ~+70°C
Temperature deviation :≤±2.0℃
Temperature fluctuation :±0.5℃
Load condition :700W
Heating Rate :≥2~3℃/min (with load 700W)
Cooling Rate :≥0.7~1℃/min (with load 700W)
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

Environmental Testing Constant Temperature Test Chamber IEC60068-2-2 512L


Complied Standards & Clauses


It conforms to standards:
IEC60068-2-2:2007 “Environmental Testing-Part 2-1: Tests- Test B-Dry heat, IDT”


Equipment Overview


Specimen Limitations:

This test equipment prohibits:

1.Testing or storage of flammable, explosive and volatile substance samples

2.Testing or storage of corrosive substance samples

3.Testing or storage of biological samples

4.Testing or storage of samples from strong electromagnetic emission sources

Test environment conditions: The ambient temperature is +25℃ and the relative humidity is ≤85%.


Technical Parameters


Technical Parameters
Control & Operation PLC and touch screen intelligent control system

Chamber Size

Inner Size External Size
800*800*800 mm (W*H*D) 1080*1950*1800 mm (W*H*D)
Temperature Range 20 °C ~+70°C
Temperature fluctuation ±0.5℃
Temperature deviation ≤±2.0℃
Heating Rate ≥2~3℃/min (with load 700W)
Cooling Rate ≥0.7~1℃/min (with load 700W)
Load condition 700W
Working noise

A sound level≤70dB(A)

(Measured in a soundproof room with an ambient temperature of 25°C and low echo. Using A weighting, test the average of 9 points. Each test point is 1 meter away from the noise source horizontally and 1 meter above the ground.)

Structural features
Insulation enclosure structure

1. Outer wall material: A3 board spraying.

2. Inner wall material: brushed stainless steel plate SUS﹟304.

3. Chamber thermal insulation material: 100mm high temperature resistant rigid polyurethane foam.

4. Chamber door insulation material: 100mm high temperature resistant rigid polyurethane foam.

Bottom structure strength The weight capacity of the track at the bottom of the test chamber: ≤100Kg/m² (load)
Air conditioning channel

1. Stainless steel long-axis centrifugal fans: 2 sets/90W.

2. Fan, heater, evaporator, drainage device, pressure balance port, adjustable air deflector, temperature sensor

Standard configuration of test chamber

1. Observation window: 400x600x40mm 3-layer vacuum tempered glass

2. Flat embedded handles

3. Door hinges: SUS #304 imported hinges

4. Energy-saving lamp inside the chamber

5. Lead hole: φ50mm 1 piece

Chamber door

1. The single-piece door opens outwards, with the hinge on the left and the handle on the right (when facing the front of the box).

2. Equipped with a safety door locking mechanism (the door can be opened in the test room), the door is equipped with electric heat to prevent condensation, and the visible range of the insulating glass observation window (W400×H600mm). The door frame is equipped with an anti-condensation electric heating device.

Control panel Controller 7-inch display screen, temperature control display screen, operating indicator light, and lighting button.
Mechanical room Refrigeration unit, compressor water tray, pressure relief device, heating device
Power distribution cabinet Distribution board, cooling fan, main power switch, transformer, intermediate relay, time relay, solid-state relay, AC contactor, circuit breaker
Heater

1. High-quality heating wire set.

2. Heating control method: SSR (Solid State Relay) non-contact equal period pulse width modulation

3. Heating power: approximately 3KW

Power cable hole Located on the back of the chamber
Details

Environmental Testing Constant Temperature Test Chamber IEC60068-2-2 512LEnvironmental Testing Constant Temperature Test Chamber IEC60068-2-2 512L

Inquiry Cart 0